SunTech CT40 Vital Signs Monitor

Product Code: N101208

Was
£695.00
Now from
£595.00 exc VAT
£714.00 inc VAT

  • SunTech Advantage™ BP Technology - Clinical-grade, market leading blood pressure monitoring
  • Manual SphygMode™ - For real-time verification of BP readings without removing the cuff and built-in memory for up to 99 charting sessions
  • Customisable BP Averaging Protocol - Ensures accurate, consistent readings (SPRINT recommended)
  • Flexibile Modular Design - Allows for the easy addition of optional Touchless IR Temperature and Nellcor SpO2 modules
  • Simple to Use - LCD display with easy to use and intuitive user interface
  • Deluxe Mobile Stand Available - Allows easy transportation of the monitor between rooms.
  • 3 year warranty

Options

Option


Related Products

SunTech Mobile Stand

Includes FREE Delivery

Was

£695.00

Now from

£595.00 exc VAT
£714.00 inc VAT

The SunTech CT40 vital signs monitor is the ideal, affordable solution for clinical-grade spot-check measurements of blood pressure and other vital signs in hospitals and clinics. 

The SunTech CT40 comes supplied with Suntech's Advantage™ BP Technology. This market leading clinical-grade blood pressure technology gives you quick and accurate monitoring of a patient's blood pressure and the Manual SphygMode™ allows for real-time verification of BP readings without the need to remove the cuff from the patient. The SPRINT recommended customisable BP averaging ensures accurate and consistent readings.

Unique to the SunTech CT40 is its modular design, which allows the easy addition or removal of optional temperature and SpO2 modules as required. Choose from Nellcor SpO2 monitoring and Touchless IR Temperature modules either at time of purchase or afterwards. The simple to use and intuitive user interface and bright LCD display ensure ease of use in the field. 

The SunTech CT40 comes supplied with Adult and Large Adult cuffs and a 3 year manufacturer's warranty. A deluxe mobile stand can be added to allow easy transportation of the monitor between rooms. A printer option is also available upon request.

Technical Specifications

Sp02 Specifications

FUNCTIONMASIMO M-LNCS DCICHIPOX
Saturation Normal (no motion) 70-100%Arms ± 2% 70-100%Arms ± 2%
Saturation Motion 70-100%Arms ± 3% N/A
Low Perfusion SpO2 +/-2% N/A
Pulse Heart rate ± 3% BPM 20-300 BPM: ±3 BPM
Pulse Heart rate w/motion ± 3% BPM 20-300 BPM: ±3 BPM

  

Temperature Specifications

FUNCTIONTOUCHLESS IR THERMOMETERCOVIDIEN ELECTRONIC THERMOMETER
Temp-erature Accuracy Body Temperature (36-39°C) ± 0.2°C Quick Mode (Oral): ± 0.3°C
Temp-erature Accuracy Body Temperature (34.0-35.9, 39.1-42.5°C) ± 0.3°C Std. Mode (Aux/Rectal): ± 0.1°C
Temp-erature Accuracy Wide Temperature (15-60°C) ± 2°C Direct Mode ± 0.1°C
Response Time less than 2 seconds 4 sec (quick mode); 10-15 sec. (Axillary/Rectal); 60 sec.(direct mode)
Resolution 0.1°C 0.1°C

  

Blood Pressure Specifications

SPECIFICATIONSVALUES
Range:
  • Pressure: Diastolic: 20-200 mmHg (adult), 20-160 mmHg (pediatric).
  • Systolic: 40-260 mmHg (adult), 40-230 mmHg (pediatric)
  • Heart Rate: 30-220 BPM (beats per minute)
Accuracy (BP): Meets or exceeds ANSI/AAMI/ISO 81060-2:2013 standard for non-invasive accuracy (±5mmHg mean error with 8mmHg standard deviation).

  

General Specifications

SPECIFICATIONSVALUES
Conditions for Use: Operating: 10°C (50°F) to 40°C (104°F) 15 – 90% RH non-condensing - 700 kPa - 1060 kPa. Operating the monitor in an environment at maximum temperature can produce temperatures exceeding 41°C (41.6°C highest recorded) on a patient-applied part. It is up to the operator to determine if this temperature is too high based upon the condition of the patient and, if so, to ensure the ambient temperature of the environment is 38°C or below.
Storage: -20°C (-4°F) to 55°C (149°F) 15 – 90% RH non-condensing - 500 kPa - 1060 kPa. Performance can be affected if used or stored outside the specified temperature, humidity, or altitude listed in the ranges above.
Power: Internal power supply. Input: 100-240 VAC @ 1.5A max, 50-60 Hz. Output +9VDC @ 5A IEC 320 type input connector.
Battery: Type: 7.2v, 6600mAh Internal Lithium-Ion battery Run Time (full charge): ~20 hrs* / 200 BP cycles Charge Time: ~6 hrs @ 23deg Celsius +/- 2deg *Unit with BP, Temp, SpO2, Wi-Fi, barcode scanner
Calibration: The accuracy of cuff -pressure transducers/indicators should be verified bi-annually.
Safety Systems: Independent hardware over-pressure circuit and redundant software overpressure algorithm to limit cuff pressure to less than 300 mmHg (+20/-10mmHg). Independent hardware timing circuit and redundant software timer algorithm to limit the duration of a blood pressure cycle to less than 180 seconds.
Dimensions: Size (without thermometer option): 8” H x 6.75’’ W x 5.2’’ D or 215mm H x 190mm W x 140mm D
Standards: ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, EN 60601-1:2006/A1:2013, IEC 60601-1-2: 2007 EMC, IEC 80601-2-30: 2013, ISO 80601-2-61: 2011, ISO 15223-1:2012, ISO 10993-1, 2009, ISO 10993-5, 2009, EN ISO 10993-10, 2010, ISO 81060-2:2013, EN 50419: 2006, EN ISO 14971:2009, CSA C22.2 No. 60601-1, EN ISO 81060-1: 2012, IEC 60601-1-6: 2013
Classifications: Equipment Classification: Class IIa per MDD, Class I (Electrical Shock), Continuous mode of operation, CE
Ingress of liquid: Ingress Protection: IPX1: This device is protected against harmful effects of dripping water per IEC 60529.